Titrimetric (Volumetric) Methods of Analysis
Term 1 - 2026
Course learning objectives
- Explain the basic principles of titrimetric analysis
- Discuss criteria for selecting primary standards
- Interpret potentiometric, thermometric and conductometric titration data
- Perform titrimetric experiments including standardization and back titration
- Perform calculations based on titrimetric data, including back titration calculations
- Cite real-world examples of titrimetric analysis
Gravimetric Methods of Analysis
Four 80-minute lessons: two theory, two practical, covering principles, equipment, execution, calculation, and application of gravimetric analysis.
L1 Β· Principles & Equipment of Gravimetric Analysis6 content Β· 0 resources
Lesson objectives
- Explain the principles of precipitation and volatilisation methods
- Describe the functions of key gravimetric equipment (crucibles, desiccators, ovens, furnaces, suction filtration apparatus)
Syllabus objectives covered
- 3.1 β explain the principles upon which gravimetric analyses are based
- 3.2 β describe the functions of the various pieces of basic equipment used in gravimetric analyses
What is Gravimetric Analysis?
Gravimetric analysis determines the quantity of a substance by mass. Instead of reading a volume off a burette, the final measurement is a mass taken on a balance - either the mass of a solid product formed from the analyte, or the change in mass of a sample after a component has been driven off.
Because the whole method rests on one accurate mass reading, technique and equipment matter enormously: every step (filtering, drying, cooling, weighing) must avoid gaining or losing mass through routes other than the reaction itself.
Two Approaches: Precipitation vs Volatilisation
Precipitation methods: the analyte is converted into an insoluble compound, which is filtered, washed, dried, and weighed. The mass of the precipitate is used, via its formula and mole ratio, to calculate how much analyte was originally present (e.g. precipitating chloride ions as AgCl, or phosphate as Cu₃(PO₄)₂).
Volatilisation methods: a sample is heated (or otherwise treated) so that a volatile component - often water - is driven off. The analyst weighs the sample before and after, and the mass lost (or mass of residue remaining) is the basis of the calculation (e.g. water of crystallisation in hydrated salts, moisture content of soils, loss-on-ignition tests).
Equipment and Their Functions
- Silica / porcelain crucible (with lid): withstands strong, repeated heating without decomposing or reacting with the sample; the lid limits spitting/loss of solid while still allowing volatiles to escape.
- Sinter-glass crucible: has a porous glass disc built into its base, allowing a precipitate to be filtered directly into the crucible under suction, then dried and weighed in the same vessel - minimising transfer losses.
- Suction flask & suction funnel: used together (often with a sinter-glass crucible or filter paper) to pull liquid through a filter under reduced pressure, speeding up filtration of a precipitate and helping it pack for washing.
- Ashless filter paper: when ordinary filter paper is used instead of a sinter-glass crucible, an 'ashless' grade is chosen so that, on ignition, it leaves behind a negligible (but not always zero) ash mass - any residual ash mass is measured with a blank and subtracted from the precipitate's mass for an accurate result.
- Oven: dries a sample or precipitate at a controlled, moderate temperature (typically well below 200°C) - used where only physically-bound moisture needs removing.
- Furnace: reaches much higher temperatures than an oven, used to ignite a precipitate to a constant, well-defined chemical form (e.g. converting a precipitate to its oxide) before final weighing.
- Desiccator: houses a drying agent and lets a hot crucible cool to room temperature without absorbing atmospheric moisture, which would add spurious mass.
Label the Gravimetric Apparatus

Precipitation or Volatilisation?
Quick Check: Principles & Equipment
L2 Β· Practical: Water of Crystallisation in Hydrated Barium Chloride5 content Β· 1 resources
Lesson objectives
- Carry out a gravimetric determination using correct technique (heating, cooling, weighing to constant mass)
- Record and organise raw experimental data for later calculation
Syllabus objectives covered
- 3.3 β perform experiments based on gravimetric determinations
- 3.4 β perform calculations based on data obtained from gravimetric analyses
Safety & Precautions
- Always use tongs to handle the hot crucible and lid - never your fingers.
- Allow the crucible to cool fully inside the desiccator before weighing; a hot object placed on a balance gives a false (drifting) reading.
- Heat gently for the first two minutes to avoid spitting of the salt, then strongly.
- Tie back hair and loose clothing near the Bunsen burner; keep the barrel collar closed until the flame is lit.
Experimental Procedure Recap
- Heat the empty crucible and lid over the Bunsen burner for 10 minutes; cool in a desiccator; weigh.
- Add approximately 2 g of hydrated barium chloride to the crucible.
- Heat gently for 2 minutes, then strongly for a further 10 minutes.
- Cool in the desiccator to room temperature; reweigh on a top-loading balance.
- Repeat the heating/cooling/weighing cycle until two consecutive masses agree (constant mass).
The Reaction on Heating
Discussion Questions
- What physical changes were observed as the barium chloride hydrate was heated?
- How did the mass of the sample change after heating?
- Why was the crucible reheated and reweighed until a constant mass was obtained?
- What is meant by the 'degree of hydration' of a salt?
- Does your experimental value for the degree of hydration match the known value? Explain any discrepancy.
- How might incomplete removal of water, or overheating, affect the calculated degree of hydration?
Quick Check
Resources
L3 Β· Calculations from Gravimetric Data & Applications in Quality Control6 content Β· 1 resources
Lesson objectives
- Convert experimental mass data into moles, mole ratios, and final quantitative answers
- Apply the calculation pathway to mixtures of two analytes
- Cite real-world quality-control applications of gravimetric analysis
Syllabus objectives covered
- 3.4 β perform calculations based on data obtained from gravimetric analyses
- 3.5 β cite examples of the use of gravimetric analysis in quality control
From Raw Masses to a Final Answer
Every gravimetric calculation follows the same pathway:
- Mass to moles - use n = m / M on the substance you actually weighed (the precipitate, or the mass lost/residue).
- Mole ratio - use the balanced equation to relate moles of the weighed substance to moles of the analyte you actually care about.
- Final answer - convert those moles of analyte into whatever the question asks for: a mass, a percentage composition, or a concentration.
Mass-Mole Relationship
Worked Example: Precipitation of Copper(II) Phosphate
Question: An excess of aqueous sodium phosphate is added to a 50.0 mL sample of copper(II) chloride solution of unknown concentration. The precipitate, Cu3(PO4)2, is filtered, dried, and weighed, giving a mass of 18.0 g.
2 Na3PO4 (aq) + 3 CuCl2 (aq) → Cu3(PO4)2 (s) + 6 NaCl (aq)
Step 1 - Molar mass of Cu3(PO4)2:
Cu: 3 × 63.55 = 190.65
P: 2 × 30.97 = 61.94
O: 8 × 16.00 = 128.00
M = 190.65 + 61.94 + 128.00 = 380.59 g/mol
Step 2 - Moles of precipitate:
n(Cu3(PO4)2) = 18.0 ÷ 380.59 = 0.0473 mol
Step 3 - Mole ratio:
From the equation, 3 mol CuCl2 produces 1 mol Cu3(PO4)2.
n(CuCl2) = 3 × 0.0473 = 0.1419 mol
Step 4 - Concentration:
Volume = 50.0 mL = 0.0500 L
[CuCl2] = 0.1419 ÷ 0.0500 = 2.84 mol dm⁻³
Challenge: Analysing a Mixture (NaCl + KCl)
Question: A 0.8870 g sample containing only NaCl and KCl is treated with excess AgNO3. The AgCl formed has a mass of 1.913 g. Calculate the %Na and %K in the sample.
Step 1 - Total moles of Cl- (= moles AgCl):
M(AgCl) = 143.32 g/mol
n(AgCl) = 1.913 ÷ 143.32 = 0.01335 mol total Cl-
Step 2 - Set up simultaneous equations.
Let x = mass NaCl, y = mass KCl, so x + y = 0.8870 g.
Moles of Cl- from each salt: x/58.44 (from NaCl) + y/74.55 (from KCl) = 0.01335
Step 3 - Solve:
Substituting x = 0.8870 - y and solving gives y (KCl) = 0.495 g and x (NaCl) = 0.392 g.
Step 4 - Convert to %Na and %K:
mass Na = 0.392 × (22.99/58.44) = 0.154 g → %Na = 0.154/0.8870 × 100 = 17.4%
mass K = 0.495 × (39.10/74.55) = 0.260 g → %K = 0.260/0.8870 × 100 = 29.3%
Note how a single precipitate mass, combined with the total sample mass, is enough to resolve a two-component mixture - as long as you set up two independent equations.
Gravimetric Analysis in Quality Control
Gravimetric analysis is used across industry and environmental monitoring wherever a precise, absolute mass-based result is needed:
- Determination of SO2 in air, wine, or fruit drinks (a food/environmental safety check).
- Determination of phosphorus (e.g. as phosphate) in fertilisers, to verify labelled nutrient content.
- Determination of chloride ion presence/concentration in water supplies, relevant to water quality standards.
Apply It: Gravimetric Calculations
Resources
L4 Β· Practical: Gravimetric Determination of Chlorides7 content Β· 0 resources
Lesson objectives
- Carry out a gravimetric determination using the precipitation method
- Apply the mass-mole-mole ratio calculation pathway to a precipitation result
- Recognise sources of error specific to precipitation gravimetric methods
Syllabus objectives covered
- 3.1 β explain the principles upon which gravimetric analyses are based
- 3.3 β perform experiments based on gravimetric determinations
- 3.4 β perform calculations based on data obtained from gravimetric analyses
Safety & Precautions
- Add dilute nitric acid to the sample before the silver nitrate - this keeps the solution acidic, preventing other insoluble silver salts (e.g. silver carbonate or silver phosphate) from precipitating alongside the AgCl.
- Add the silver nitrate in slight excess to ensure all the chloride present is precipitated.
- Protect the precipitate from strong/direct light where possible - AgCl is light-sensitive and can slowly decompose, which would affect the final mass.
- Wash the precipitate thoroughly with distilled water to remove soluble impurities, but avoid excessive washing, which can dissolve a small amount of the AgCl itself.
Experimental Procedure Recap
- Weigh a sample of the solid metal chloride.
- Dissolve the sample in water.
- Add nitric acid, then excess silver nitrate, to the chloride solution.
- The chloride precipitates as silver chloride.
- Collect the precipitate by filtration.
- Wash the precipitate with distilled water.
- Dry the precipitate.
- Weigh the mass of silver chloride formed.
The Precipitation Reaction
Worked Example: Determining %Cl in an Impure Sample
Question: In the analysis of 0.7011 g of an impure chloride-containing sample, 0.9805 g of AgCl was precipitated. What is the percentage by mass of chloride in the sample?
Step 1 - Molar mass of AgCl:
M(AgCl) = 107.87 (Ag) + 35.45 (Cl) = 143.32 g/mol
Step 2 - Moles of precipitate (= moles Cl-, 1:1 ratio):
n(AgCl) = 0.9805 ÷ 143.32 = 0.006842 mol
Step 3 - Mass of chloride:
mass Cl = 0.006842 × 35.45 = 0.2425 g
Step 4 - Percentage by mass:
%Cl = 0.2425 ÷ 0.7011 × 100 = 34.6%
Extension: Correcting for a Filter-Paper Blank
Question: A 1.1105 g sample of bauxite was analysed for aluminium. The precipitate was ignited on filter paper to form pure Al2O3, giving a total mass of 0.3605 g. A blank test showed that 10 identical sheets of filter paper leave an average ash mass of 0.0006 g each. What is the percentage aluminium in the bauxite?
Step 1 - Correct the precipitate mass for the paper's own ash:
corrected mass Al2O3 = 0.3605 - 0.0006 = 0.3599 g
Step 2 - Moles of Al2O3:
M(Al2O3) = (2 × 26.98) + (3 × 16.00) = 101.96 g/mol
n(Al2O3) = 0.3599 ÷ 101.96 = 0.003530 mol
Step 3 - Moles and mass of Al:
n(Al) = 2 × 0.003530 = 0.007060 mol
mass Al = 0.007060 × 26.98 = 0.1905 g
Step 4 - Percentage aluminium:
%Al = 0.1905 ÷ 1.1105 × 100 = 17.2%
This is exactly why an ashless filter paper (Lesson 1) or a blank correction matters - without it, this result would be too high.